Operation steps of leather material tensile testing machine

Aug 03, 2021 Leave a message

1. Make the sample: Make the sample according to the requirements of the test standard, and measure its size and other data.

2. Power on: first turn on the software measurement and control system on the computer, and then turn on the power of the testing machine. Select the appropriate test type according to the test to be performed.

3. Trial operation: Select the appropriate speed to make the testing machine run up and down to make sure that the tensile machine is running normally. 4. New record: According to the number of tests to be performed, create a new number of corresponding test records; and fill in the corresponding batch number, serial number, test environment, sample size and other relevant data.

5. Load the sample: adjust the position of the sample on the testing machine, and load the sample.

6. Select the range: according to the test force and deformation range required for the test, adjust the tension machine to the appropriate test force and deformation gear (range).

7. Reset: reset the test force, deformation and displacement to zero.

8. Test speed: According to the requirements of the test process in the standard, set an appropriate test speed; if there is no speed requirement in the standard, set a more appropriate speed, and the speed should not be too large, so as not to affect the test results.

9. End condition: According to the requirements of the test process in the standard, if the sample is to be destroyed (such as tensile, cement compression), select "Break judgment"; if the test force is pulled (or compressed), it will end. , select "Target" and "Test Force"; if it ends by pulling (or pressing) to a certain deformation, select "Target" and "Deformation".

10. Test start: Click the "Test Start" button to start the test.

11. The end of the test: After the test is completed, the "test end" button is automatically pressed. If you need to end the test halfway, you can also press the "Test End" button to end the test manually.

12. Test results: Click the "Analyze" button to view the corresponding test results. Continue the test: Click the "Test" button to return to the test operation interface, click the "Next" button to switch the test record to the next one, and repeat steps 5-11 to test.

13. Save data: If the test data needs to be saved, press the "Save" button to save the test data to the database for reference. Print report: If the test results need to be printed, press the "Output Report" button to output the test results to a Word document and print.

14. Shutdown: Turn off the power of the testing machine first, then turn off the software measurement and control system and computer.